Sirin Diagnostics HCG Pregnancy Test Cassette

Sirin Diagnostics – HCG Pregnancy Test (Cassette)

Rapid test for the detection of HCG in serum or urine

Intended Use

The HCG Pregnancy Test is a rapid, visual, and qualitative immunoassay for the detection of human chorionic gonadotropin (HCG) in serum or urine specimens, to aid in the early detection of pregnancy.

  • Manufactured under ISO 13485:2016
  • Made in Egypt by Sirin Diagnostics
  • Bulk, distributor & private-label supply

Product Details

Principle
The test is based on a chromatographic immunoassay principle. The strip contains:
Test line (T): coated with anti-alpha HCG antibodies to capture HCG.
Control line (C): coated with anti-mouse antibodies to verify proper test performance.
If HCG is present in the specimen, a pink line will appear in the test region along with the control line. The control line must always appear to validate the result.
—
Kit Components
Individually packaged test device.
Buffer solution.
Instructions for use.
—
Storage
Store at 2–30°C.
Do not freeze.

Stable until the expiration date printed on the pouch.

—
Specimen Collection

Serum: Separate from whole blood and refrigerate (2–8°C) for up to 3 days or freeze for long-term storage.

Urine: Collect in a clean, dry container. First morning urine is preferred for higher HCG concentration.
—

Test Procedure

1. Bring the test device and specimen to room temperature before use.

2. Open the pouch and place the device on a flat surface.

3. Add 3 drops of serum or urine into the sample well.

4. Wait 5–8 minutes for lines to appear. Do not read after 10 minutes.

—

Interpretation of Results

Positive: Two pink lines appear (T and C) → HCG detected (pregnant).

Negative: One pink line at C only.

Invalid: No control line (C) appears.
—
Performance Characteristics

Analytical sensitivity: 25 mIU/ml or less.

Diagnostic sensitivity: 100%.

Diagnostic specificity: 100%.

No cross-reaction with LH, FSH, or TSH at specified concentrations.
—
Limitations

Avoid excessive fluid intake before testing.

Certain tumors or HCG-containing medications may cause false positives.

Very early pregnancies may not be detected; retest after 48 hours.

A confirmed diagnosis of pregnancy should only be made by a physician.

Specifications

Catalog No.Available on request
Product lineWomen's Health
Intended useProfessional in-vitro diagnostic use
FormatCassette
Specimen typeSee IFU
Result timeSee IFU
StorageSee IFU / label
Shelf lifeAvailable on request
Pack sizeAvailable on request
ManufacturerSirin Diagnostics, Egypt (ISO 13485:2016)

For professional in-vitro diagnostic use only. Need full technical data? Request product information.